Software & SaMD classification

Regulatory basis

MDR software classification is governed by Annex VIII, Rule 11 of MDR 2017/745. IVDR software follows the same class assignment rules as other IVDs under IVDR Annex VIII. MDCG 2021-24 provides authoritative guidance on MDR Rule 11 application.

Disclaimer

This site provides general information only and does not constitute legal or regulatory advice. Software qualification and classification requires careful analysis — consult MDCG guidance and a qualified regulatory professional.


Step 1 — Qualification: is it a medical device at all?

Before classifying software, establish that it qualifies as a medical device under MDR Art. 2(1) or as an IVD under IVDR Art. 2(2). Software qualifies only if it:

  1. Is intended by the manufacturer for a medical purpose, and
  2. Performs an action that goes beyond storage, archival, lossless compression, or simple search of data

Software that merely stores or transmits data without interpretation, calculation, or decision support does not qualify as a medical device. Examples:

SoftwareQualification
PACS (picture archiving and communication system) — storage onlyNot a medical device
AI algorithm reading ECG and flagging arrhythmiasMedical device (MDR)
EHR system — administrative onlyNot a medical device
Clinical decision support giving individualised treatment recommendationMedical device (MDR)
Laboratory LIMS — routing samples, no interpretationNot an IVD
Algorithm interpreting blood gas values and producing diagnosisIVD (IVDR)

Step 2 — MDR Rule 11 classification

Once qualified as an MDR medical device, software is classified under Annex VIII Rule 11 using a tiered approach based on the severity of the condition and reversibility of the impact of an erroneous output.

Rule 11 class tiers

Class III — software intended to provide information used to make decisions with diagnosis or therapy purposes where those decisions could cause:

  • Death, or
  • Irreversible deterioration of a person's state of health

Class IIb — software intended to provide information used to make decisions with diagnosis or therapy purposes where those decisions could cause:

  • Serious deterioration of a person's state of health (reversible), or
  • Surgical intervention

Class IIa — software intended to:

  • Provide information used to make decisions with diagnosis or therapy purposes other than the above, or
  • Monitor physiological processes

Class I — all other software (e.g. software intended for general health and wellness purposes without a specific medical claim that crosses the qualification threshold)

MDCG decision tree (MDCG 2021-24)

The MDCG 2021-24 guidance provides a formal decision tree. The key branches are:

  1. Is the software driving or influencing clinical decision-making?
  2. What is the condition it addresses — life-threatening, serious, or minor?
  3. Is the impact of an error reversible with timely intervention?
Condition / ImpactClass
Life-threatening conditions; errors cause death or irreversible harmIII
Serious conditions; errors cause reversible serious harm or require surgical interventionIIb
Non-serious conditions; errors cause minor harm or no clinical intervention requiredIIa
General wellness / non-diagnostic / non-therapeuticI (if qualified as MD at all)

Worked examples — MDR Rule 11

SoftwareReasoningClass
AI triage tool flagging potential stroke on CTDiagnosis of life-threatening condition; delay = irreversible harmIII
Automated insulin dosing algorithm (closed-loop pump)Therapy for chronic condition; overdose → serious reversible harmIIb
Spirometry interpretation software (flags possible COPD)Diagnosis of serious condition; error → further testing, not immediate harmIIa
BMI calculator in a wellness appNo specific medical diagnostic or therapeutic claimI (if MD)
Radiation therapy planning systemTherapy directly affecting delivery of ionising radiationIIb or III depending on configuration
AI ECG algorithm flagging AF for GP reviewDiagnosis of significant arrhythmia; actionable recommendationIIb
Patient-reported outcome (PRO) questionnaire with clinical scoringMonitoring tool providing clinical informationIIa

Step 3 — IVDR software classification

Software that qualifies as an IVD under IVDR is classified under the standard IVDR Rules 1–7, not Rule 11. The same risk-based logic applies — what information does the software provide, and what is the consequence of an error?

Examples:

  • Algorithm interpreting genomic sequencing to identify cancer mutations → IVDR Class C or D depending on clinical use
  • Software calculating eGFR from creatinine result → supports clinical chemistry output → Class B typically
  • AI pathology platform reading histology slides for cancer diagnosis → IVDR Class C

Key concepts for software classification

Intended purpose is everything

Rule 11 classification depends entirely on what the manufacturer states the software is for — not how it might be used. A manufacturer choosing a conservative intended purpose (e.g. "for decision support only, not for diagnosis") may lower the classification, but must ensure all promotional materials, labelling, and IFU are consistent with that stated purpose.

Off-the-shelf (OTS) software components

Software components that are general-purpose and not specifically intended for medical use do not require MDR/IVDR compliance on their own. However, if they are incorporated into a medical device software system, the overall system must comply. The manufacturer of the integrated device is responsible for ensuring OTS components do not compromise compliance.

AI/ML-specific considerations

AI and machine learning software in medical devices raises additional questions:

  • Adaptive algorithms that change with use may require re-evaluation if the change goes beyond the scope of what was assessed
  • Locked algorithms are easier to document and validate
  • The MDCG and European Commission are developing further guidance on AI in medical devices (aligned with the EU AI Act, which has separate obligations for high-risk AI systems — many of which overlap with medical device AI)

Software lifecycle (EN 62304)

Software classified as Class IIa, IIb, or III must follow a documented software development lifecycle. The harmonised standard EN 62304:2006+A1:2015 (Medical device software — software life cycle processes) defines:

  • Software safety classification (Class A, B, C under 62304)
  • Development, verification, and maintenance requirements proportionate to safety class


Official references

ReferenceDescription
MDR Annex VIII, Rule 11Software classification rule
MDR Art. 2(1)Definition including software
IVDR Art. 2(2)IVD definition including software
MDCG 2019-11Qualification and classification of software
MDCG 2021-24MDR Rule 11 classification guidance
EN 62304:2006+A1:2015Software lifecycle processes
EU AI Act (Regulation (EU) 2024/1689)Parallel obligations for AI systems