Labelling Requirements
TR-007 ยท HK-unique bilingual requirement
Minimum Label Contentsโ
- Device name/trade name
- Manufacturer name and address
- LRP name and address (HK contact)
- Lot/batch number or serial number
- Manufacturing date and/or expiry date
- Intended purpose (if not obvious)
- Warnings and precautions
- Single-use symbol (if applicable)
Bilingual Requirement โ English & Chineseโ
HK-unique: Devices intended for home use or lay users must be labelled in both English and Chinese.
Professional-use devices (used only by trained clinicians) may use English or Chinese only, but bilingual is always safer for market acceptance.
IFU (Instructions for Use)โ
- Must accompany the device unless the label alone is sufficient
- Home-use IFU must be bilingual
- Electronic IFU (eIFU) may be acceptable for certain professional-use devices
Promotional Claimsโ
Must be truthful, not misleading, and consistent with the MDACS-listed intended purpose.