Skip to main content

Biocompatibility (ISO 10993)

For devices with direct or indirect patient contact, MFDS requires a biocompatibility evaluation per the ISO 10993 series.

Key standardsโ€‹

StandardScope
ISO 10993-1Evaluation and testing within risk management
ISO 10993-5In vitro cytotoxicity
ISO 10993-10Skin sensitization
ISO 10993-11Systemic toxicity
ISO 10993-18Chemical characterization

Contact categories (ISO 10993-1)โ€‹

Surface contacting ยท External communicating ยท Implant โ€” combined with contact duration (transient < 24h / short-term โ‰ค 30 days / long-term > 30 days) determines required tests.

A biocompatibility evaluation matrix must be included in the STED. Tests must be conducted by a recognised testing laboratory.