Skip to main content

How classification works in New Zealand

Risk-based classification

New Zealand classifies medical devices based on their potential risk to patients and users. Higher-risk devices are subject to closer regulatory scrutiny and require stronger conformity evidence. Lower-risk devices have lighter requirements.

Classification is set out in Schedule 2 of the Medicines (Database of Medical Devices) Regulations 2003 — a set of 22 rules covering non-IVD devices. IVD classification follows separate rules.


Who classifies the device?

Classification is the sponsor's (and manufacturer's) responsibility

Medsafe does not classify devices. Classification is the responsibility of the manufacturer or sponsor, and must be justified in technical documentation. Medsafe does not offer pre-market classification rulings. If you are uncertain, consult the manufacturer or seek independent regulatory advice.


The classification hierarchy — non-IVDs

New Zealand uses five classes for non-IVD devices, in increasing order of risk:

ClassRisk levelExamples
Class ILowestBandages, examination gloves, walking frames
Class IIaLow-mediumBlood transfusion sets, ultrasound equipment, sutures
Class IIbMedium-highVentilators, surgical lasers, bone screws
Class IIIHighestHeart valves, coronary stents, implantable contraceptives
AIMDSpecific high-riskPacemakers, cochlear implants, implantable defibrillators

The classification hierarchy — IVDs

IVDs use a separate set of classes (lower-case numbers):

ClassRisk levelExamples
Class 1LowestGeneral laboratory instruments, specimen collection items
Class 2Low-mediumBlood glucose self-tests, pregnancy tests
Class 3Medium-highBlood grouping systems, HIV screening tests
Class 4HighestCJD screening, blood group confirmatory tests
IVDs and WAND

IVDs are exempt from mandatory WAND notification under Schedule 1 of the Regulations. They can be voluntarily notified, and many sponsors choose to do so for documentation purposes.


How the 22 rules work

The 22 classification rules in Schedule 2 are applied based on the device's:

  • Intended purpose — what the manufacturer says the device is for
  • Duration of contact — transient (≤60 min), short-term (≤30 days), long-term (>30 days)
  • Degree of invasiveness — non-invasive, invasive, surgically invasive, implantable
  • Anatomical location — body surface, body orifice, vascular, cardiac, central nervous system
  • Energy use — whether the device emits or transfers energy, and of what type

If more than one rule applies, the highest resulting class takes precedence.

See The 22 Classification Rules for rule-by-rule guidance.


NZ vs EU and AU classification

New Zealand's rules are based on the original GHTF (Global Harmonization Task Force) framework — the same foundation used by the EU Medical Devices Directive and the Australian TGA prior to the EU MDR transition.

Because New Zealand has not adopted the EU MDR (2017/745) classification changes, some devices that moved from Class IIa to IIb (or from IIb to III) under the EU MDR may still carry their older, lower NZ classification. Always verify against Schedule 2 — never assume your EU MDR class is your NZ class.