Skip to main content

FSCA Process

Step 1 — Internal FSCA Decision

The manufacturer makes the internal decision to initiate an FSCA. Document the rationale, scope (affected devices, lot numbers, serial numbers), and the chosen corrective action type.

Step 2 — Swissmedic Notification

The manufacturer or CH REP must notify Swissmedic before or simultaneously with the start of the FSCA. Notification is via the Swissmedic eVigilance portal and must include: description of the issue; devices affected (UDI, lot/serial numbers); corrective action planned; draft Field Safety Notice (FSN); timeline for implementation.

Step 3 — Field Safety Notice (FSN)

The FSN is the written communication sent to all customers affected by the FSCA. It must:

  • Be clearly titled "URGENT FIELD SAFETY NOTICE" or equivalent
  • Describe the issue and the risk to patients/users
  • Clearly identify the affected devices (UDI, lot numbers, model names)
  • Specify the required action for the customer (e.g. stop use, return, retrofit, advisory)
  • Include contact details for questions and for returning devices (if applicable)
  • Be submitted to Swissmedic before or simultaneously with dispatch to customers
  • Be in German, French, and Italian for Swiss customers

Step 4 — Customer Notification and Implementation

Send the FSN to all identified customers (hospitals, distributors, healthcare professionals). Maintain a record of all notifications sent and acknowledgements received. Implement the corrective action (device return, retrofit, etc.).

Step 5 — Effectiveness Check

After implementation, conduct an effectiveness check to confirm the corrective action has been received and acted upon by customers. Document the results and report to Swissmedic in the follow-up report.

Official Sources

Disclaimer

AI-assisted content for navigation only. Always verify against official Swissmedic and Fedlex sources. Not legal or regulatory advice.