Skip to main content

MDR vs IVDR — which applies?

note

Regulatory basis - MDR = Regulation (EU) 2017/745 | IVDR = Regulation (EU) 2017/746. Both regulations are mutually exclusive — a product is governed by one or the other, never both simultaneously.

warning

Disclaimer - This site provides general information only and does not constitute legal or regulatory advice. Always consult the official regulation text and a qualified regulatory professional.


The fundamental split

RegulationCovers
MDR 2017/745Medical devices, accessories to medical devices, Annex XVI non-medical-purpose devices
IVDR 2017/746In vitro diagnostic medical devices and their accessories

The split is determined entirely by the product's intended purpose and mode of use: is the examination performed in vitro on a human-derived specimen, or is the device used in some other way?


The primary test

Ask this question first:

Is the device intended to examine a specimen derived from the human body, outside the body, to provide diagnostic or monitoring information?

AnswerRegulation
Yes — in vitro examination of human-derived specimensIVDR 2017/746
NoMDR 2017/745 (if it is a medical device at all)

Decision flowchart

Is the product intended for use in human beings?

├─ NO → Not an MDR/IVDR medical device

└─ YES → Does it act principally by pharmacological/immunological/metabolic means?

├─ YES → Medicinal product (Directive 2001/83/EC)

└─ NO → Is it used in vitro on human-derived specimens
to provide diagnostic/monitoring information?

├─ YES → IVDR 2017/746

└─ NO → MDR 2017/745

Common scenarios

Clearly MDR

ProductReason
Hip implantImplanted in the body
Insulin pumpDelivers therapy in the body
Cardiac monitor (wearable)Measures in vivo physiological signals
MRI scannerImaging performed on the body
Surgical robotUsed on the body
Dermal fillerInjected into the body (Annex XVI)
Contact lens (cosmetic)Used on the body (Annex XVI)
Wound dressingApplied to the body
SaMD for radiology image analysisAnalyses in vivo imaging data

Clearly IVDR

ProductReason
Blood glucose test stripIn vitro examination of blood specimen
PCR assay for infectious diseaseIn vitro examination of patient sample
BRCA genetic testIn vitro examination of tissue/blood
Blood typing reagentIn vitro compatibility testing
Urine pregnancy testIn vitro examination of urine specimen
Clinical chemistry analyserProcesses in vitro blood/urine samples
Companion diagnosticIdentifies patients from specimen
Histopathology analysis softwareExamines tissue specimens in vitro

Borderline — requires careful analysis

ProductIssueResolution
Point-of-care blood glucose meterIn vitro, but used at bedsideIVDR — near-patient testing is still in vitro
Continuous glucose monitor (CGM)Sensor worn on/in body; no specimen extractionMDR — in vivo interstitial measurement
NGS sequencing platformGeneral-purpose instrument with IVD assay kitsIVDR for the IVD-intended configuration as a system
Smart toilet analysing urineWellness vs diagnostic claimsIf diagnostic disease claims: IVDR; wellness only: likely neither
Microbiome analysis kitExamines stool/salivaIf disease diagnosis/monitoring claims: IVDR

Key pathway differences

TopicMDRIVDR
ClassificationClass I · IIa · IIb · III (Rules 1–22)Class A · B · C · D (Rules 1–7)
Notified body required?No for Class I (except Is/Im/Ir); Yes for IIa+No for Class A; Yes for most B, C, D
Clinical/performance evidenceClinical evaluation (Art. 61, Annex XIV)Performance evaluation (Art. 56, Annex XIII)
Post-market follow-upPMCFPMPF
PSUR requiredClass IIb, IIIClass C, D
Companion diagnostic rulesN/AClass C minimum; EMA consultation required

When you are still unsure

Consult the MDCG Borderline and Classification Working Group guidance. Key documents:

  • MDCG 2022-5 — Borderline manual between MDR and other legislation
  • MDCG 2019-11 — Software qualification and classification

National competent authorities can provide formal opinions. These are not binding EU-wide but are a recognised route to regulatory certainty.



Official references

ReferenceDescription
MDR Art. 2(1)Medical device definition
IVDR Art. 2(2)IVD definition
MDR Art. 1(8)Interaction with medicinal product rules
MDCG 2022-5Borderline product guidance
MDCG 2019-11Software qualification guidance