Skip to main content

What is an in vitro diagnostic medical device?

note

Regulatory basis - This page is based on Article 2(2) of Regulation (EU) 2017/746 (IVDR). The IVD definition determines whether a product is governed by IVDR rather than MDR.

warning

Disclaimer - This site provides general information only and does not constitute legal or regulatory advice. Always consult the official regulation text and, where appropriate, a qualified regulatory professional.


Under Article 2(2) of the IVDR, an in vitro diagnostic medical device (IVD) is any medical device that is a reagent, reagent product, calibrator, control material, kit, instrument, apparatus, piece of equipment, software, or system — used alone or in combination — and that is:

  1. Intended by the manufacturer for in vitro use (outside the living body), and
  2. Applied to specimens derived from the human body (including blood, tissue, organ donations, etc.), and
  3. Intended solely or principally to provide information on one or more of the following:
Information typeExamples
Physiological or pathological process or stateBlood glucose level, haemoglobin count, inflammatory markers
Congenital physical or mental impairmentsNIPT for trisomy, newborn screening panels
Predisposition to a medical condition or diseaseBRCA1/2 genetic testing, HLA typing
Safety and compatibility with potential recipientsBlood group typing, cross-matching, donor compatibility
Treatment response or reactionsPharmacogenomics, companion diagnostic testing
Defining or monitoring therapeutic measuresTherapeutic drug monitoring, INR for anticoagulation

Specimen receptacles

Specimen receptacles — containers or materials specifically intended to receive and preserve specimens derived from the human body for diagnostic examination — are also regulated as IVDs. This includes vacutainers, specimen tubes, and urine cups designed to preserve specimen integrity.

Accessories to IVDs

An accessory to an IVD is an article intended by its manufacturer to be used together with one or more specific IVDs, specifically enabling or assisting the IVD to be used according to its intended purpose, or solely or primarily assisting the safety or performance of the IVD.


The "in vitro" requirement

"In vitro" means the examination takes place outside the living human body, on a specimen that has been derived from it. This distinguishes IVDs from:

  • In vivo devices (e.g. implantable sensors, endoscopes) → these are medical devices under MDR
  • In vivo diagnostic imaging (e.g. MRI, ultrasound) → MDR

The specimen must be derived from a human being. Tests performed exclusively on animal specimens are not within IVDR scope (they fall under veterinary legislation).


What IVDR covers that MDR does not

IVDR introduces several elements not present under the old IVDD 98/79/EC:

AreaChange under IVDR
Classification4 risk classes (A, B, C, D) replacing the old List A / List B / General system
Notified body involvementNow required for most Class B, C, and D devices (previously only List A and B)
Performance evaluationStructured, lifecycle approach — not just analytical performance
EUDAMEDRegistration and UDI obligations now apply to IVDs
PMS and PMPFFormal post-market performance follow-up required
Common specificationsEC can issue CSs for specific IVD categories

Software as an IVD

Software is explicitly included in the IVDR definition. IVD software qualification follows the same general logic as MDR software, but with IVD-specific criteria:

  • Does the software examine specimens derived from the human body?
  • Is the output intended to provide diagnostic or monitoring information from those specimens?

Key examples:

  • An algorithm that interprets sequencing data from a tissue biopsy to identify mutations → IVD software
  • A laboratory information management system (LIMS) that routes samples but does not interpret results → likely not an IVD
  • AI-based image analysis of histopathology slides → may be borderline — assess against intended purpose carefully

See the MDCG 2019-11 guidance on software qualification for detailed worked examples.


Companion diagnostics

A companion diagnostic (CDx) is an IVD device essential for the safe and effective use of a corresponding medicinal product. Under IVDR:

  • CDx devices are always classified as Class C (at minimum)
  • The conformity assessment procedure requires coordination with the competent authority evaluating the medicinal product (EMA or national)
  • The notified body is required to seek a scientific opinion from EMA or a national medicines authority as part of the conformity assessment

This creates a mandatory interaction between the IVDR pathway and the pharmaceutical regulatory pathway. See Companion diagnostics for a detailed walkthrough.


The "intended purpose" principle

As with MDR, the IVDR definition turns on manufacturer intent. The intended purpose is established through:

  • Labelling and instructions for use
  • Promotional materials
  • Technical documentation
  • Any other communication from the manufacturer about what the device is for

A laboratory instrument sold as a general-purpose photometer is not an IVD. The same instrument sold with a claim to measure haemoglobin concentration in blood is an IVD.


Key questions to ask

  1. Is the examination in vitro? The test must be performed on a specimen outside the body.
  2. Is the specimen from a human being? Veterinary IVDs are out of scope.
  3. What information does it provide? Does the output fit one of the six categories listed in Art. 2(2)?
  4. Is it software? If so, does it interpret specimen-derived data, or merely process/store it?
  5. Is it also used on the body? If a device has both in vitro and in vivo functions, assess each function separately.
  6. Could it be MDR instead? See MDR vs IVDR — which applies?.


Official references

ReferenceDescription
IVDR Art. 2(2)Definition of IVD
IVDR Art. 2(4)Definition of accessory to an IVD
IVDR Art. 2(12)Definition of intended purpose
IVDR Art. 48(3)Companion diagnostic conformity assessment
IVDR Annex VIIIClassification rules for IVDs
MDCG 2019-11Qualification and classification of software
MDCG 2020-16Guidance on companion diagnostics