Class IV — Specially Controlled Medical Devices (Highest Risk)
Class IV is the highest-risk category of medical device under Japan's PMD Act classification system. All Class IV devices must use the Shonin (full approval) pathway — there is no Ninsho option.
Characteristics
Class IV devices:
- Are implanted in the human body for more than 30 days, OR
- Contact the central nervous system, cardiovascular system, or vital organs and are life-sustaining/life-supporting, OR
- Are IVDs used for blood transfusion compatibility or infectious disease screening (e.g. HIV)
- If defective, pose a risk of death or serious irreversible injury
Examples of Class IV devices
Drug-eluting stents, pacemakers, ICDs (implantable cardioverter defibrillators), artificial heart valves, cochlear implants, HIV diagnostic IVDs, SARS-CoV-2 IVDs used for blood screening, total hip and knee replacements with Class IV designation.
Regulatory pathway — Shonin only
Class IV devices always require full Shonin approval. Key features:
- Full PMDA scientific review with typically 18–24+ month standard review timelines
- Comprehensive clinical evidence package required — foreign clinical data may be accepted with appropriate justification
- QMS inspection of all manufacturing sites (both domestic and foreign) as part of the approval process
- Conditions on approval are common — e.g. requirements for post-market registry studies
Re-examination period
Class IV devices are subject to a re-examination period of typically 4–7 years (varies by device type), during which extensive post-market use-results data must be collected and submitted to MHLW.