Technical Dossier Requirements
The technical dossier (expediente técnico) is the package of documentation submitted to COFEPRIS to support a sanitary registration application under the standard route. Requirements scale with device class.
Core dossier elements
1. Device description & intended use
- Generic name and trade name of the device.
- Clear description of the device, its components, materials, and accessories.
- Intended use statement — the clinical purpose, patient population, and conditions of use.
- Contraindications (if applicable).
2. Technical specifications
- Design drawings or schematics (as applicable).
- Material specifications.
- Dimensional specifications.
- Sterilisation method and validation data (for sterile devices).
- Shelf life and storage conditions.
3. Biocompatibility
For devices that contact the patient, biocompatibility data in accordance with ISO 10993 is required:
- List of materials in contact with the patient.
- Biocompatibility testing results or justification for test exemption.
4. Clinical evidence
| Class | Clinical evidence expectation |
|---|---|
| Class I | Literature review; may use published data on substantially similar devices |
| Class II | Literature review + clinical evaluation report; clinical study data for novel or higher-risk Class II |
| Class III | Comprehensive clinical evidence including clinical study data, performance evaluation, benefit–risk analysis |
5. Risk management
A risk management summary or report prepared in accordance with ISO 14971, covering:
- Identified hazards and hazardous situations.
- Risk control measures.
- Residual risk assessment.
6. GMP documentation
Evidence of quality system compliance — one of the following:
- ISO 13485 certificate (most widely accepted).
- MDSAP audit report.
- Certificate of Good Manufacturing Practices issued by a recognised authority.
- Evidence of NOM-241-SSA1-2025 compliance.
7. Labelling
- Draft Spanish-language label compliant with NOM-137-SSA1-2008.
- Instructions for Use (IFU) in Spanish.
- Sterilisation and storage symbols (compliant with ISO 15223).
8. Administrative documents
- Power of attorney appointing the MRH (for foreign manufacturers).
- Proof of fee payment (pago de derechos).
- MRH's operating notice (Aviso de funcionamiento).
IVD-specific additions
For IVD devices, include:
- Analytical performance data (sensitivity, specificity, precision, linearity).
- Clinical performance validation data.
- Reference intervals and calibration information.
- Stability data for reagents.
Language requirement
All documents submitted to COFEPRIS must be in Spanish or accompanied by an officially certified Spanish translation.