Class I self-declaration pathway
Class I self-declaration is governed by Schedule 3, Annex VII equivalent of Part I of the UK MDR 2002. Class I devices with sterile, measuring, or reusable surgical instrument aspects require partial UKAB involvement under the relevant annexes of Schedule 3.
This site provides general information only and does not constitute legal or regulatory advice. Always verify your device's classification before relying on the self-declaration route.
Who can use the self-declaration pathway?โ
The self-declaration pathway is available for standard Class I medical devices โ those that are:
- Not placed on the market in a sterile condition
- Do not have a measuring function
- Are not reusable surgical instruments
If your Class I device falls into any of those three sub-categories, partial UKAB involvement is required (see below).
The self-declaration processโ
Step 1: Prepare the technical fileโ
The manufacturer must compile a technical file demonstrating conformity with all applicable Essential Requirements (Schedule 1, UK MDR 2002). The technical file must include:
- General description of the device and its intended purpose
- Classification rationale (confirming Class I status under Schedule 2)
- Essential Requirements checklist with evidence of conformity for each applicable ER
- Design specifications, drawings, and manufacturing information
- Risk management documentation (ISO 14971 risk file)
- Clinical evaluation report or literature-based clinical evidence
- Labelling and instructions for use (in English for the GB market)
- Labelling samples or artwork
- Quality management system description (ISO 13485 or equivalent)
The technical file does not need to be submitted to MHRA proactively โ but it must be available for inspection on request, and must be retained for a minimum of 5 years from the date the last device was manufactured (or 15 years for implantable devices, noting most Class I devices are not implantable).
Step 2: Issue the UK Declaration of Conformityโ
The manufacturer issues a UK Declaration of Conformity (UK DoC) โ a formal written declaration that:
- Identifies the manufacturer and (if applicable) the UKRP
- Describes the device (name, model, intended purpose, device identifier)
- States that the device meets the Essential Requirements of the UK MDR 2002
- References the Schedule 3 annex used (Annex VII equivalent for Class I self-declaration)
- Is signed and dated by a responsible person at the manufacturer
The UK DoC must be kept on file and is available to MHRA and economic operators on request.
Step 3: Affix the UKCA markโ
The manufacturer affixes the UKCA mark to the device (or its packaging/IFU where the device itself is too small or impractical to mark). The UKCA mark must appear clearly and legibly. For Class I self-declaration, the UKCA mark appears alone โ no UKAB identification number is added.
Step 4: Register with MHRAโ
Before placing the device on the GB market, the manufacturer (or their UKRP for non-UK manufacturers) must register via the MHRA Submissions portal.
Registration includes:
- Manufacturer/UKRP legal entity details
- Device description and classification
- Declaration of Conformity reference
Step 5: Place on the GB marketโ
Once steps 1โ4 are complete, the device may be placed on the GB market. Post-market obligations (PMS, vigilance reporting) begin at this point.
Class I sub-categories: where UKAB involvement is requiredโ
Class I (sterile) โ devices placed on the market in a sterile conditionโ
The manufacturer self-declares all Essential Requirements except those relating to sterility. For sterility aspects, a UKAB must:
- Audit the sterilisation process validation and QMS
- Issue a certificate covering the sterilisation process
The UKCA mark on sterile Class I devices is accompanied by the UKAB's identification number.
Documentation required beyond standard Class I technical file:
- Sterilisation validation records (ISO 11135, 11137, 11138, or equivalent depending on sterilisation method)
- Sterile barrier system validation (ISO 11607)
- Environmental monitoring records for cleanroom manufacture
- Bioburden data
Class I (measuring function) โ devices with a measuring functionโ
A measuring function means the device has a scale or numeric output where the accuracy of the measurement directly affects clinical decisions (e.g., a graduated syringe, a non-electronic thermometer).
The manufacturer self-declares all ERs except those relating to the measuring function. A UKAB must assess the metrology aspects โ verifying that the device meets accuracy and precision requirements relevant to its measuring function.
The UKCA mark on measuring Class I devices is accompanied by the UKAB's identification number.
Class I (reusable surgical instruments โ Class Im)โ
Reusable surgical instruments โ instruments intended by the manufacturer to be reprocessed and used multiple times โ require UKAB assessment of the aspects relating to reprocessing and reuse information. The UKAB reviews whether:
- The manufacturer provides adequate instructions for cleaning, disinfection, sterilisation, and inspection between uses
- The device is designed to withstand the specified number of reprocessing cycles
The UKCA mark on Class Im devices is accompanied by the UKAB's identification number.
CE marking and transitional acceptanceโ
Class I devices from non-UK manufacturers currently relying on CE marking (under EU MDD or EU MDR) can continue to place those devices on the GB market under the transitional arrangements until 30 June 2030, without obtaining UKCA marking.
However, MHRA registration remains mandatory regardless of whether the device bears CE or UKCA marking.
Common pitfalls for Class I self-declarationโ
| Pitfall | How to avoid |
|---|---|
| Claiming Class I for a device that should be Class IIa | Apply Schedule 2 rules rigorously; document rationale |
| Technical file too thin | Class I does not mean minimal evidence โ all ERs must be addressed |
| Forgetting MHRA registration | Registration is mandatory even for self-declared Class I devices |
| No clinical evaluation | Even Class I devices require clinical evidence; literature review may suffice |
| Missing UKAB for sterile/measuring/Im sub-categories | Check whether your Class I device falls into a sub-category |
| Inadequate IFU | IFU must comply with Schedule 1 ER 18 regardless of class |
Related pagesโ
- Conformity assessment โ overview
- Class IIa ยท IIb ยท III โ UK Approved Body routes
- UKCA marking requirements
- MHRA registration
- Technical documentation overview
- Essential Requirements overview
Official referencesโ
| Reference | Description |
|---|---|
| UK MDR 2002, Schedule 3, Annex VII equivalent | Class I declaration of conformity procedure |
| UK MDR 2002, Schedule 3 | Conformity assessment procedures for all classes |
| MHRA: Conformity assessment for medical devices | MHRA guidance on conformity assessment routes |
| ISO 13485:2016 | QMS standard |
| ISO 11607:2019 | Packaging for terminally sterilised medical devices |